IS09001 IS014001 / IS045001 certification enterprises

Durability performance test

YT3600 geocell oxidation induction period analyzer:

About our product:

This machine measures the relationship between temperature and heat flow related to the internal thermal transformation of the material. It has a wide range of applications, especially the research and development, performance testing and quality control of materials. Material characteristics: such as glass transition temperature.
  • Product details
【Scope of application】
This machine measures the relationship between temperature and heat flow related to the internal thermal transformation of the material. It has a wide range of applications, especially the research and development, performance testing and quality control of materials. Material characteristics: such as glass transition temperature.
【Standards compliant】
Q/CR549-2016 Railway Engineering Geosynthetics Technical Standard
GB/T 19466.1-2004 Plastic Differential Scanning Calorimetry (DSC) Part 1: General Rules
GB/T 19466.2-2004 Plastic Differential Scanning Calorimetry (DSC) Part 2: Determination of Glass Transition Temperature
GB/T 19466.3-2004 Plastic Differential Scanning Calorimetry (DSC) Part 3: Determination of melting and crystallization temperature and enthalpy
GB/T 19466.6-2009 Plastic Differential Scanning Calorimetry (DSC) Part 6: Determination of Oxidation Induction Time (Isothermal OIT) and Oxidation Induction Temperature (Dynamic OIT)
【main feature】
1. Brand new furnace structure, better resolution and resolution, and better baseline stability
2. Digital gas mass flowmeter, accurately control the purge gas flow, and the data is directly recorded in the database
3. The instrument can adopt two-way control (host control, software control), with a friendly interface and easy operation.
【technical parameter】
1. DSC range: 0~±600m
2. Temperature range: room temperature~600℃ air cooling
3. Heating rate: 0.1~100℃/min
4. Temperature resolution: 0.1℃
5. Temperature fluctuation: ±0.1℃
6. Temperature repeatability: ±0.1℃
7. DSC noise: 0.01μW
8. DSC resolution: 0.01μW
9. DSC accuracy: 0.1μW
10. DSC sensitivity: 0.001mW
11. Temperature control method: heating, constant temperature (full program automatic control)
12. Curve scan: heating scan
13. Atmosphere control: automatic switching of the instrument
14. Display mode: 24bit color, 7-inch LCD touch screen display
15. Data interface: standard USB interface
16. Parameter standard: equipped with standard material (tin), users can calibrate the temperature and enthalpy by themselves.
17. Appearance size: 500*400*155mm (length, width and height)
Previous     Next